FDA Recall Open, Classified

Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System

Recall: Z-1049-2023 · Initiated December 16, 2022

Recall

Recall Number
Z-1049-2023
Event Number
91467
Firm
NovaSignal Corp.
FEI Number
3013368530
Product Code
IYN
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
December 16, 2022
Posted
February 1, 2023
Address
2440 S Sepulveda Blvd, Los Angeles, CA, 90064-1784

Description

Lucid TCD 2.0 System, Model: NA-LUCM1. Transcranial Ultrasound System

Reason

Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.

Action

On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive. Complete and return the customer response form. Customers with questions can contact Customer Operations at [email protected]

Distribution

US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL

Quantity

6