FDA Recall Terminated

LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes

Recall: Z-1048-2019 · Initiated February 6, 2019

Recall

Recall Number
Z-1048-2019
Event Number
82214
Firm
LW Scientific, Inc.
FEI Number
3000251573
Product Code
JQC
Status
Terminated
Root Cause
Device Design
Initiated
February 6, 2019
Terminated
May 7, 2021
Address
865 Marathon Pkwy, Lawrenceville, GA, 30046-2857

Description

LW Scientific ZIP IQ PCV Centrifuge, ZIC-24HD-75T3 Rev C Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes

Reason

The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.

Action

The recall was initiated by telephone on 02/06/2019. Initial requesting that the consignee immediately cease use of the product. Recall notices were disseminated by telephone and email requesting the return of the product beginning 02/13/2019. The recall is to the user level. The firm has posted a notice on their website stating there is a voluntary recall due to a safety issue and asking users to call if they have a unit purchased after 9/28/18.

Distribution

Worldwide Distribution - US Nationwide and the countries of Philippines, Honduras, El Salvador, Australia, Guatemala, Canada, Singapore, Australia, Peru, Nicaragua, Trinidad and Tobago, and New Zealand

Quantity

142 units