FDA Recall Terminated

ABBOTT TestPack+Plus Strep A with On Board Controls (OBC) II, list numbers 5C63-16 (20 test kit) and 5C63-21(40 test kit); Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064 USA

Recall: Z-1048-04 · Initiated June 11, 2004

Recall

Recall Number
Z-1048-04
Event Number
29352
Firm
Abbott Laboratories HPD/ADD/GPRD
FEI Number
1415939
Product Code
GTZ
Status
Terminated
Root Cause
Other
Initiated
June 11, 2004
Posted
July 20, 2004
Terminated
October 5, 2004
Address
100/200 Abbott Park Road, Abbott Park, IL, 60064

Description

ABBOTT TestPack+Plus Strep A with On Board Controls (OBC) II, list numbers 5C63-16 (20 test kit) and 5C63-21(40 test kit); Abbott Laboratories, Diagnostics Division, Abbott Park, IL 60064 USA

Reason

False Positive results may be obtained with the identified lots when using a negative patient sample or a negative control.

Action

A Device Recall Letter was sent to customers that received the affected product lots on 6/11/04 via Federal Express Priority. The accounts were informed of the false positive readings, and were requested to destroy any affected inventory of the recalled lots.

Distribution

The product was not distributed in the U.S. The product was distributed internationally through Abbott affiliates in Argentina, Korea, the United kingdom, Japan, Singapore, the Dominican Republic, Canada, Hong Kong, Chile and Germany

Quantity

75,368 kits