FDA Recall Terminated

Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemodialysis treatment.

Recall: Z-1047-2015 · Initiated December 19, 2014

Recall

Recall Number
Z-1047-2015
Event Number
70288
Firm
Fresenius Medical Care Holdings, Inc.
FEI Number
3001451489
Product Code
KOC
Status
Terminated
Root Cause
Device Design
Initiated
December 19, 2014
Posted
February 3, 2015
Terminated
February 18, 2021
Address
920 Winter St, Waltham, MA, 02451-1521

Description

Fresenius Crit Line in a Clip (CLiC) with SW version 2.51 Model Number: CL10041001. A continuous real-time monitor for non-invasive hematocrit, oxygen saturation and percent change in blood volume calculation during hemodialysis treatment.

Reason

Potential for misinterpretation of the graphic display of the Blood Volume ( BV) slope

Action

FMCRTG, LLC representatives contacted Clinics on 12/19/14 by visit/telephone phone) and removed all CLiC units.

Distribution

CT, NY

Quantity

35 units