FDA Recall Open, Classified

Optima XR646 HD. X-Ray imaging system.

Recall: Z-1046-2025 · Initiated January 28, 2025

Recall

Recall Number
Z-1046-2025
Event Number
96208
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
KPR
Status
Open, Classified
Root Cause
Radiation Control for Health and Safety Act
Initiated
January 28, 2025
Posted
February 3, 2025
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Optima XR646 HD. X-Ray imaging system.

Reason

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

Action

Firm sent an "IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING" email/letter to its consignees. The email/letter described the product, the problem and the actions to be taken. The consignees were instructed to do the following: You can continue to use the device as intended for clinical use, by following the instructions below: 1.If you perform an exposure and observe an on-screen message that the AEC termination limit has been reached, please acknowledge the message. 2.Take any relevant actions to adjust techniques or patient positioning for the next exposure to avoid reaching the termination limit for that exposure, if appropriate. Please ensure that all potential users in your facility are made aware of this notification and the recommended actions. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.

Quantity

3149 (1697 US; 1452 OUS)