FDA Recall Open, Classified

SMV DSI, System, Tomography, Computed, Emission

Recall: Z-1045-2026 · Initiated December 10, 2025

Recall

Recall Number
Z-1045-2026
Event Number
98133
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel
FEI Number
3004571294
Product Code
KPS
Status
Open, Classified
Root Cause
Package design/selection
Initiated
December 10, 2025
Posted
January 9, 2026

Description

SMV DSI, System, Tomography, Computed, Emission

Reason

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.

Action

GE Healthcare began notifying their consignees by telephone on 12/10/2025, followed by an URGENT MEDICAL DEVICE CORRECTION UPDATE notice distributed on 01/05/2025 via US mail. The notice explained the issue, safety risk, and requested the consignee stop use of the product. Care should be exercised when deinstalling these systems. To discuss replacement options, your GE HealthCare account representative. Should be contacted.

Distribution

US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece

Quantity

3 units