Ambra PACS UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R Software Version Number: V3.16.13.0 Intended for use as a primary diagnostic and analysis tool for diagnostic images.
Recall
- Recall Number
- Z-1045-2017
- Event Number
- 76187
- Firm
- DICOM GRID INC
- FEI Number
- 3008776294
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software change control
- Initiated
- December 16, 2016
- Posted
- January 18, 2017
- Terminated
- April 7, 2017
- Address
- 4530 E Muirwood Dr, Ste. 113, Phoenix, AZ, 85048-7639
Description
Ambra PACS UDI: +AMBRHEALTHSOLUTIONS0/$$+7_V3.16.13.0R Software Version Number: V3.16.13.0 Intended for use as a primary diagnostic and analysis tool for diagnostic images.
A software error caused the window/level to become the same in one series regardless if the image had different levels; image results have a washed-out grey appearance.
On 12/16/16 a fix is the software issue was conducted via remote cloud system. On 01/11/17 customers were sent Release Notes that addressed the software correction. The issue was resolved and required no further action.
U.S. distribution to the following; TX, FL, NE, CA, WI, AZ, NH, MO, MA, CO, PA, NV, PR, IL, OH, NY, MD, NJ, MN, MI, CT, VA, GA, UT, KS, WA, MI, HI. Foreign distribution to the following; Canada, Spain, Israel, Singapore, Kosovo, Austria, Brazil, Malaysia, Australia.
209