The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models. Vascular Embolization Device Prestige Coil Model Numbers and Product description: PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack PRES0930H
Recall
- Recall Number
- Z-1042-2024
- Event Number
- 93751
- Firm
- BALT USA, LLC
- FEI Number
- 3014162263
- Product Code
- KRD
- Status
- Open, Classified
- Root Cause
- Material/Component Contamination
- Initiated
- January 3, 2024
- Posted
- February 8, 2024
- Address
- 29 Parker, Ste 100, Irvine, CA, 92618-1667
Description
The Prestige Coil System is a product family of embolic coils with associated delivery system components. These devices are angiographically delivered through the vasculature to embolize peripheral vascular abnormalities. The devices are to be placed to create blood stasis, reducing flow into the anomaly, and thrombosing the target site. The Prestige Coil System is comprised of the implant coil and a pusher assembly. There are a total of 81 Prestige coil models. Vascular Embolization Device Prestige Coil Model Numbers and Product description: PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES1137COM18 Prestige Coil 11mm x 37cm Complex-18 PRES0202CXPPL Prestige Coil 2mm x 2cm Prestige Plus Complex Xtra Pac PRES0203CXPPL Prestige Coil 2mm x 3cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES0315HXPPL Prestige Coil 3mm x 15cm Prestige Plus Helical Xtra Pac PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0620HPKPL Prestige Coil 6mm x 20cm Prestige Plus Helical Pack PRES0720CPKPL Prestige Coil 7mm x 20cm Prestige Plus Complex Pack PRES0930CPKPL Prestige Coil 9mm x 30cm Prestige Plus Complex Pack PRES0101HXPPL Prestige Coil 1mm x 1cm Prestige Plus Helical Xtra Pac PRES0153HXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Helical Xtra Pac PRES0720HPKPL Prestige Coil 7mm x 20cm Prestige Plus Helical Pack PRES0830HPKPL Prestige Coil 8mm x 30cm Prestige Plus Helical Pack PRES0930HPKPL Prestige Coil 9mm x 30cm Prestige Plus Helical Pack PRES1040HPKPL Prestige Coil 10mm x 40cm Prestige Plus Helical Pack PRES1240COM18 Prestige Coil 12mm x 40cm Complex-18 PRES0204CXPPL Prestige Coil 2mm x 4cm Prestige Plus Complex Xtra Pac PRES0206CXPPL Prestige Coil 2mm x 6cm Prestige Plus Complex Xtra Pac PRES0310CXPPL Prestige Coil 3mm x 10cm Prestige Plus Complex Xtra Pac PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0830CPKPL Prestige Coil 8mm x 30cm Prestige Plus Complex Pack PRES0210CXP10 Prestige Coil 2mm x 10cm Complex-10 Xtra Pack PRES0258CXP10 Prestige Coil 2.5mm x 8cm Complex-10 Xtra Pack PRES0417CPK10 Prestige Coil 4mm x 17cm Complex-10 Xtra Pack PRES0520CPK10 Prestige Coil 5mm x 20cm Complex-10 Xtra Pack PRES0615CPK10 Prestige Coil 6mm x 15cm Complex-10 Xtra Pack PRES0103CXPPL Prestige Coil 1mm x 3cm Prestige Plus Complex Xtra Pac PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CPKPL Prestige Coil 4mm x 13cm Prestige Plus Complex Pack PRES2525HXPPL Prestige Coil 2.5mm x 2.5cm Prestige Plus Helical Xtra Pac PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0307CXPPL Prestige Coil 3mm x 7cm Prestige Plus Complex Xtra Pac PRES0517CPKPL Prestige Coil 5mm x 17cm Prestige Plus Complex Pack PRES0620CPKPL Prestige Coil 6mm x 20cm Prestige Plus Complex Pack PRES0152CXPPL Prestige Coil 1.5mm x 2cm Prestige Plus Complex Xtra Pac PRES0413CRG10 Prestige Coil 4mm x 13cm Complex 10 Regular PRES0153CXPPL Prestige Coil 1.5mm x 3cm Prestige Plus Complex Xtra Pac PRES0255CXPPL Prestige Coil 2.5mm x 5cm Prestige Plus Complex Xtra Pac PRES0358CXPPL Prestige Coil 3.5mm x 8cm Prestige Plus Complex Xtra Pac PRES0410CPKPL Prestige Coil 4mm x 10cm Prestige Plus Complex Pack PRES1040CPKPL Prestige Coil 10mm x 40cm Prestige Plus Complex Pack PRES0420HPKPL Prestige Coil 4mm x 20cm Prestige Plus Helical Pack PRES0520HPKPL Prestige Coil 5mm x 20cm Prestige Plus Helical Pack PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0206HXP10 Prestige Coil 2mm x 6cm Helical-10 Xtra Pack PRES1447COM18 Prestige Coil 14mm x 47cm Complex-18 PRES0204HXP10 Prestige Coil 2mm x 4cm Helical-10 Xtra Pack PRES0306HPK10 Prestige Coil 3mm x 6cm Helical-10 Pack PRES0310HPK10 Prestige Coil 3mm x 10cm Helical-10 Pack PRES0410HPK10 Prestige Coil 4mm x 10cm Helical-10 Pack PRES0930H
Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.
On 01/03/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via priority mail to customers informing them that Balt USA has became aware of a situation where discoloration along the delivery system pusher component of the Optima Coil System was identified during in-house inspection of released finished good products. The discoloration was observed on the outside of the PET jacket along the hypotube component. Customers are instructed to: 1) Stop or cease use of the affected recall product. 2) Inform their hospital staff, such as safety officers, pharmacists, purchasing agent, head of neuroradiology and the neuroradiology staff, physicians including but not limited to endovascular neurosurgeons and interventional neurologists, as well as any other person if deemed necessary of the recall. 3) Identify and locate the Optima Coil System products concerned by this recall procedure. 4) Collect and put in quarantine the Optima Coil System products concerned by this recall procedure, then return them to Balt USA through the usual RGA (Return Goods Authorization) procedure by contacting Balt USA s Customer Service department. 5) Keep informed by Balt USA about the status of every unit of Optima Coil System products concerned by this recall procedure. For questions or additional assistance, contact Balt USA Quality Department at: Contact: Quality Department email: [email protected] Balt USA, LLC 29 Parker | Irvine CA, 92618 | USA Telephone: +1 949.788.1443 | Fax: +1 949.788.1444
U.S.: AZ, CA, CO, FL, GA, HI, IL, KS, MA, ME, MO, MS, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI and Puerto Rico O.U.S.: Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Ecuador, France, Indonesia, India, Japan, Korea, Malaysia, Mexico, Panama, Paraguay, Peru, Russia, Uruguay, and Vietnam.
2802 coils