FDA Recall Terminated

Leica M822 Surgical Microscope Device Controller Software (MDC version 3.6, pack F, affecting the control unit component, Article #10448446, in combination with a Leica M822 Optics Carrier-XY-Focus-Tilt-Unit, article #10448587; Leica Microsystems (Schweiz) AG, Max Schmidheiny-Strasse 201, CH-9425, Heerbrugg, Switzerland The intended use of the microscope is for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Recall: Z-1042-2012 · Initiated December 21, 2011

Recall

Recall Number
Z-1042-2012
Event Number
60866
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
ETP
Status
Terminated
Root Cause
Device Design
Initiated
December 21, 2011
Posted
February 16, 2012
Terminated
February 16, 2012
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

Leica M822 Surgical Microscope Device Controller Software (MDC version 3.6, pack F, affecting the control unit component, Article #10448446, in combination with a Leica M822 Optics Carrier-XY-Focus-Tilt-Unit, article #10448587; Leica Microsystems (Schweiz) AG, Max Schmidheiny-Strasse 201, CH-9425, Heerbrugg, Switzerland The intended use of the microscope is for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Reason

Leica Microsystems received complaints stating that the zoom function of the Leica M822 surgical microscope locked up during use.

Action

The firm, Leica Microsystems, Inc., sent a "Medical Device Correction" notice dated December 21, 2011 to all affected customers by e-mail. The product, problem, and actions to be taken by the customers were addressed. Customers were instructed- DO NOT USE use the affected product until the software could be upgraded and complete and return the Acknowledgement Form via fax to: 1-847-607-3147 or Scan and send to [email protected]. Leica Microsystems representatives will contact the customers to make necessary arrangements to complete the software upgrade. If you have any questions, contact Director, Global Product Quality Management at 1-847-317-7209.

Distribution

Worldwide distribution: USA (nationwide) including states of: California, Colorado, Illinois, New York, Oregon, South Carolina; and countries of: Australia, Japan, Korea, Pakistan, Romania, South Africa, Switzerland and Turkey.

Quantity

29 units (10 units-USA)