FDA Recall Terminated

Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by Bio-Rad Laboratories, Hercules, CA Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood. This device only tests for A 1 c.

Recall: Z-1041-2009 · Initiated October 17, 2008

Recall

Recall Number
Z-1041-2009
Event Number
50174
Firm
Bio-Rad Laboratories Inc
FEI Number
2915274
Product Code
LCP
Status
Terminated
Root Cause
Pending
Initiated
October 17, 2008
Posted
March 24, 2009
Terminated
November 12, 2010
Address
4000 Alfred Nobel Dr, Hercules, CA, 94547-1803

Description

Bio-Rad Variant II TURBO Hemoglobin Testing System with CDM v 4.0, model number 270-2600, Manufactured by Bio-Rad Laboratories, Hercules, CA Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A 1 c in EDTA human whole blood. This device only tests for A 1 c.

Reason

Sample results may be assigned to the wrong barcode number and tube position numbers, leading to discrepant sample results.

Action

An Urgent Medical Device Correction letter was sent out on October 17, 2008 via fax and phone call. The letter recommended that customers review the Summary Report for duplicate tube position numbers or barcode IDs within the same rack. If there are questions or a need for assistance, customers are to contact their regional Bio-Rad office. Follow-ups will be made once root cause is identified.

Distribution

Worldwide Distribution

Quantity

92 units