FDA Recall Terminated

SHIMADZU MobileDaRt Evolution (MX8 Version), Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR).

Recall: Z-1039-2019 · Initiated January 14, 2019

Recall

Recall Number
Z-1039-2019
Event Number
81940
Firm
Shimadzu Medical Systems Usa Com
FEI Number
1000125362
Product Code
IZL
Status
Terminated
Root Cause
Device Design
Initiated
January 14, 2019
Terminated
April 13, 2023
Address
20101 S Vermont Ave, Torrance, CA, 90502-1328

Description

SHIMADZU MobileDaRt Evolution (MX8 Version), Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR).

Reason

In these units, the brake is usually released by gripping the drive handle and applied by letting go of the handle. However, for systems within a specified manufacture period, we have identified that even when letting go of the handle, the systems brakes are not applied. In this event, if the Operator releases the handle when the device is moving, the unit will continue to coast.

Action

Shimadzu Medical Systems sent an Urgent Medical Device Correction Notice letter dated January 7, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The corrective action for systems identified with this issue is replacement of the systems drive handle. Customers will be contacted by a Shimadzu or Fuji local Service Representative to determine whether your system requires the above correction action. For questions contact the Shimadzu National Service Support group at (844) 487-2767.

Distribution

US Nationwide Distribution

Quantity

93 units