SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research or clinical laboratory.
Recall
- Recall Number
- Z-1038-2015
- Event Number
- 70021
- Firm
- SCC Soft Computer
- FEI Number
- 1000307651
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- May 20, 2014
- Posted
- February 2, 2015
- Terminated
- May 16, 2016
- Address
- 5400 Tech Data Drive, Clearwater, FL, 33760
Description
SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research or clinical laboratory.
The interface fails to send abnormal flags for Reference Lab test results.
The firm sent out 2 CORRECTION COMMUNICATIONS informing consignees of functionality in release of 4.0.7 HIS Add-on interfaces which SCC has designated to be a potential patient Risk-to-Health issue. Further the corrective action indicated HIS releases 4.0.7.2 and higher will be corrected. Hot Fixes have been developed for HIS Add-on patches 4.0.7.0 and 4.0.7.1. Lastly the communication requests that the user "please acknowledge receipt of this task and grant permission to load the mandatory hot fix(es). Thank you."
Distributed US (nationwide) in the states of OH, GA, NJ, WI, FL, NH, IL, NY, IN, CA, AL, VA, MO, MI, KY, MA, PA, CO, ME, TN, WA, MN, SD, LA, NC, and NV and the country of Canada.