FDA Recall Terminated

SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research or clinical laboratory.

Recall: Z-1038-2015 · Initiated May 20, 2014

Recall

Recall Number
Z-1038-2015
Event Number
70021
Firm
SCC Soft Computer
FEI Number
1000307651
Product Code
JQP
Status
Terminated
Root Cause
Device Design
Initiated
May 20, 2014
Posted
February 2, 2015
Terminated
May 16, 2016
Address
5400 Tech Data Drive, Clearwater, FL, 33760

Description

SoftLab with SA HIS versions 4.0.7.0-4.0.7.1 SoftLab is a laboratory information system to be used in a medical research or clinical laboratory.

Reason

The interface fails to send abnormal flags for Reference Lab test results.

Action

The firm sent out 2 CORRECTION COMMUNICATIONS informing consignees of functionality in release of 4.0.7 HIS Add-on interfaces which SCC has designated to be a potential patient Risk-to-Health issue. Further the corrective action indicated HIS releases 4.0.7.2 and higher will be corrected. Hot Fixes have been developed for HIS Add-on patches 4.0.7.0 and 4.0.7.1. Lastly the communication requests that the user "please acknowledge receipt of this task and grant permission to load the mandatory hot fix(es). Thank you."

Distribution

Distributed US (nationwide) in the states of OH, GA, NJ, WI, FL, NH, IL, NY, IN, CA, AL, VA, MO, MI, KY, MA, PA, CO, ME, TN, WA, MN, SD, LA, NC, and NV and the country of Canada.

Quantity

48