FDA Recall Open, Classified

1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack

Recall: Z-1037-2022 · Initiated March 22, 2022

Recall

Recall Number
Z-1037-2022
Event Number
89868
Firm
American Contract Systems, Inc.
FEI Number
3003419114
Product Code
OJH
Status
Open, Classified
Root Cause
Process control
Initiated
March 22, 2022
Address
2610 Ne Industrial Dr, Ste 220, Kansas City, MO, 64117-2648

Description

1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack

Reason

Product was sterilized with a higher than specification EO concentration.

Action

A customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following: 1) Review your inventory and segregate affected product. This product should not be shipped out to customers. 2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120. 3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness. A Field Action Response Form should be completed and returned to [email protected], even if you do not have product inventory.

Distribution

US Nationwide distribution in the states of IL, MO, NE, and TX.

Quantity

114 packs