FDA Recall Terminated

Diapact CRRT Assembled Kit HF/HD with Manifold and Standard Spike, item number 7210517. The product is shipped 4 units per shipping carton. Hemodialysis kit, B. Braun Medical, Inc., Allentown, PA 18109

Recall: Z-1036-2008 · Initiated December 10, 2007

Recall

Recall Number
Z-1036-2008
Event Number
46348
Firm
B. Braun Medical, Inc.
FEI Number
2523676
Product Code
FJK
Status
Terminated
Root Cause
Process control
Initiated
December 10, 2007
Posted
May 7, 2008
Terminated
October 28, 2008
Address
901 Marcon Boulevard, Allentown, PA, 18109-9512

Description

Diapact CRRT Assembled Kit HF/HD with Manifold and Standard Spike, item number 7210517. The product is shipped 4 units per shipping carton. Hemodialysis kit, B. Braun Medical, Inc., Allentown, PA 18109

Reason

Faulty tubing does not prime machine as intended.

Action

B. Braun Medical sent an Urgent Medical Device Reall letter dated 12/12/07. Customers were told to discontinue use of the tubing.

Distribution

Nationwide Distribution.

Quantity

956 units