FDA Recall
Terminated
Diapact CRRT Assembled Kit HF/HD with Manifold and Standard Spike, item number 7210513. The product is shipped 4 units per shipping carton. Hemodialysis kit, B. Braun Medical, Inc., Allentown, PA 18109
Recall: Z-1035-2008
·
Initiated December 10, 2007
Recall
- Recall Number
- Z-1035-2008
- Event Number
- 46348
- Firm
- B. Braun Medical, Inc.
- FEI Number
- 2523676
- Product Code
- FJK
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- December 10, 2007
- Posted
- May 7, 2008
- Terminated
- October 28, 2008
- Address
- 901 Marcon Boulevard, Allentown, PA, 18109-9512
Description
Diapact CRRT Assembled Kit HF/HD with Manifold and Standard Spike, item number 7210513. The product is shipped 4 units per shipping carton. Hemodialysis kit, B. Braun Medical, Inc., Allentown, PA 18109
Reason
Faulty tubing does not prime machine as intended.
Action
B. Braun Medical sent an Urgent Medical Device Reall letter dated 12/12/07. Customers were told to discontinue use of the tubing.
Distribution
Nationwide Distribution.
Quantity
1544 units