FDA Recall Terminated

Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.

Recall: Z-1032-04 · Initiated May 20, 2004

Recall

Recall Number
Z-1032-04
Event Number
29228
Firm
Sechrist Industries Inc
FEI Number
2020676
Product Code
CBK
Status
Terminated
Root Cause
Other
Initiated
May 20, 2004
Posted
July 20, 2004
Terminated
March 14, 2006
Address
4225 E La Palma Ave, Anaheim, CA, 92807-1815

Description

Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.

Reason

Electronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all.

Action

Recall notification contains a test method to be used to identify defective units. Return notification that ventilator is defective will result in replacement of the electronic manometer.

Distribution

Nationwide and to Malaysia, Japan, Kuwait, Sri Lanka, South Africa, Guatemala, Mexico, Morocco, Indonesia, Korea, Ecuador, Colombia, Portugal, Oman, Switzerland, Spain, Zimbabwe, Belgium, Brazil, Thailand, Argentina, Costa Rica, Taiwan, ROC, India, Russia, Saudi Arabia, Pakistan, Caribbean, Venezuela.

Quantity

240