Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.
Recall: Z-1032-04
·
Initiated May 20, 2004
Recall
- Recall Number
- Z-1032-04
- Event Number
- 29228
- Firm
- Sechrist Industries Inc
- FEI Number
- 2020676
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- May 20, 2004
- Posted
- July 20, 2004
- Terminated
- March 14, 2006
- Address
- 4225 E La Palma Ave, Anaheim, CA, 92807-1815
Description
Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.
Reason
Electronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all.
Action
Recall notification contains a test method to be used to identify defective units. Return notification that ventilator is defective will result in replacement of the electronic manometer.
Distribution
Nationwide and to Malaysia, Japan, Kuwait, Sri Lanka, South Africa, Guatemala, Mexico, Morocco, Indonesia, Korea, Ecuador, Colombia, Portugal, Oman, Switzerland, Spain, Zimbabwe, Belgium, Brazil, Thailand, Argentina, Costa Rica, Taiwan, ROC, India, Russia, Saudi Arabia, Pakistan, Caribbean, Venezuela.
Quantity
240