FDA Recall
Terminated
Depuy brand LCS Complete metal backed patella, cemented, LRG+; REF 1294-09-770.
Recall: Z-1031-05
·
Initiated June 7, 2005
Recall
- Recall Number
- Z-1031-05
- Event Number
- 32282
- Firm
- Depuy Orthopaedics, Inc.
- FEI Number
- 1818910
- Product Code
- NJL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 7, 2005
- Posted
- July 21, 2005
- Terminated
- May 2, 2006
- Address
- 700 Orthopaedic Dr, Warsaw, IN, 46582-3994
Description
Depuy brand LCS Complete metal backed patella, cemented, LRG+; REF 1294-09-770.
Reason
There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.
Action
Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.
Distribution
Nationwide and international.