FDA Recall Terminated

AxSYM Folate Reagent Pack; list 03C81 ; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064

Recall: Z-1029-2007 · Initiated April 23, 2007

Recall

Recall Number
Z-1029-2007
Event Number
37887
Firm
Abbott Laboratories
FEI Number
1415939
Product Code
CGN
Status
Terminated
Root Cause
Other
Initiated
April 23, 2007
Posted
July 11, 2007
Terminated
May 18, 2008
Address
100 Abbott Park Rd, Abbott Park, IL, 60064-3502

Description

AxSYM Folate Reagent Pack; list 03C81 ; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064

Reason

Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.

Action

Abbott initiated a field correction recall by letter dated 4/23/07 to all customers receiving the affected reagents. The accounts were informed of the failure of the AxSYM to open (actuate) some reagent packs with the affected lots. The accounts were requested to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading the reagent pack on the instrument; and discard any reagent packs that do not open properly. If the user observes a probe crash, prior to testing samples again, the user must perform a probe crash recovery following the procedure in the Operations Manual. If necessary, they need to replace the probe. If a probe crash recovery is not performed, subsequent results may be impacted. The customers were requested to retain a copy of the letter for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.

Distribution

Worldwide, including USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Hong Kong, Mexico, New Zealand, Singapore, South Korea, Taiwan, Thailand and Venezuela.

Quantity

1,711 packs