FDA Recall Terminated

Evans Wedge Disposable Trials, Large Footprint, 8mm Thickness (Model: 6101-22200818). Orthopedic manual surgical instrument

Recall: Z-1028-2022 · Initiated March 30, 2022

Recall

Recall Number
Z-1028-2022
Event Number
89983
FEI Number
3014833750
Product Code
LXH
Status
Terminated
Root Cause
Process change control
Initiated
March 30, 2022
Posted
April 29, 2022
Terminated
August 28, 2023
Address
311 W Corporation St, Inc. restor3d, Durham, NC, 27701-2409

Description

Evans Wedge Disposable Trials, Large Footprint, 8mm Thickness (Model: 6101-22200818). Orthopedic manual surgical instrument

Reason

During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.

Action

An email was sent out to the consignees on 03/31/2022. All customers were instructed to segregate any affected product and contact Restor3d. If all product has already been returned to restor3d, the consignee is to indicate that on the enclosed response form. On 04/19/2022, the firm extended the recall to two additional lots. The recall notification was sent to the consignees via email with a request for all affected product to be returned.

Distribution

US Nationwide distribution in the states of KY, NC, NJ, TX, IN, VA and AL.

Quantity

9 units