Fresenius 2008T Series Hemodialysis System
Recall
- Recall Number
- Z-1026-2017
- Event Number
- 76037
- Firm
- Fresenius Medical Care Renal Therapies Group, LLC
- FEI Number
- 3001451489
- Product Code
- KDI
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 21, 2016
- Terminated
- March 17, 2023
- Address
- 920 Winter St, Waltham, MA, 02451-1521
Description
Fresenius 2008T Series Hemodialysis System
When the UF Rate, Goal or Time is adjusted using the up and down arrow keys, and the change is cancelled by using the esc key, the cancelled UF Rate is actually being executed rather than rate displayed on the machine
Fresenius issued Customer Notification via Certified Mail on 12/21/2016. Users provided with problem and recommendation, when adjusting the UF Rate, UF Goal, or Tx Time, is to use the numeric data entry keys instead of the up or down arrow keys. Change the values by using data entry keys, then save by pressing CONFIRM. Fresenius will make arrangements to install software on the hemodialysis machine. Confirmation of the noticed by the consignee will be required via a Reply Form.
Nationwide Foreign: Canada, Mexico, Jamaica, Germany, Guam
65361 units (US) 249 units (OUS)