FDA Recall Open, Classified

ADVIA Centaur Enhanced Estradiol (eE2) (REF) 500T - In vitro diagnostic use in the quantitative detection of estradiol SMN: 10697757 All lots

Recall: Z-1025-2022 · Initiated March 3, 2022

Recall

Recall Number
Z-1025-2022
Event Number
89953
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
1219913
Product Code
CHP
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 3, 2022
Address
333 Coney St, East Walpole, MA, 02032-1516

Description

ADVIA Centaur Enhanced Estradiol (eE2) (REF) 500T - In vitro diagnostic use in the quantitative detection of estradiol SMN: 10697757 All lots

Reason

Falsely Elevated Results Observed with Plasma Specimens, could lead a clinician to misinterpret patient results

Action

Siemens issued an Urgent Medical Device Correction AIMC 22-03.A.US and Urgent Field Safety Notice AIMC 22-03.A.OUS on 3/3/22. Letter states reason, health risk and action to take: Customers must discontinue use of plasma specimens for testing with the Atellica IM and ADVIA Centaur eE2 assays until further notice. Customers may continue to use the Atellica IM eE2 and ADVIA Centaur eE2 assays with serum specimens. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Questions: contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, Bulgaria, Burkina, Faso Country AG, Cyprus, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Greece, Guadeloupe, India, Ireland, Israel, Italy, Kazakhstan, Latvia, Lithuania, Macedonia, Malta, Morocco, Nepal, Netherlands, Norway, Palestinian Ter, Paraguay, Poland, Portugal, Qatar, Romania, Russian Fed., Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., Ukraine, United Kingdom, Vatican city.

Quantity

88 units