Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.
Recall
- Recall Number
- Z-1024-2013
- Event Number
- 64448
- Firm
- Zyno Medical LLC
- FEI Number
- 3006575795
- Product Code
- OPP
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- June 13, 2011
- Posted
- March 29, 2013
- Terminated
- July 25, 2016
- Address
- 10 Tech Circle, Natick, MA, 01760
Description
Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.
Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump
Zyno Medical initiated recall 6/13/11 to accounts advising of a Preventative Maintenance (PM) Schedule and a software upgrade Odometer to monitor counts of accumulative fluid volume infused by the unit. On 3/1/13, firm issued correction notice to users that have not had the software implemented on their unit. Units will be serviced at their scheduled PM. The notice identified the product, problem, and actions to be taken by the customers.
Nationwide Distribution.
2589 units