FDA Recall Terminated

Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.

Recall: Z-1024-2013 · Initiated June 13, 2011

Recall

Recall Number
Z-1024-2013
Event Number
64448
Firm
Zyno Medical LLC
FEI Number
3006575795
Product Code
OPP
Status
Terminated
Root Cause
Process design
Initiated
June 13, 2011
Posted
March 29, 2013
Terminated
July 25, 2016
Address
10 Tech Circle, Natick, MA, 01760

Description

Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.

Reason

Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump

Action

Zyno Medical initiated recall 6/13/11 to accounts advising of a Preventative Maintenance (PM) Schedule and a software upgrade Odometer to monitor counts of accumulative fluid volume infused by the unit. On 3/1/13, firm issued correction notice to users that have not had the software implemented on their unit. Units will be serviced at their scheduled PM. The notice identified the product, problem, and actions to be taken by the customers.

Distribution

Nationwide Distribution.

Quantity

2589 units