FDA Recall Open, Classified

FreeStyle Libre 3 Plus Sensor REF 72080-01 UDI-DI code: 00357599819002 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.

Recall: Z-1023-2026 · Initiated November 24, 2025

Recall

Recall Number
Z-1023-2026
Event Number
98049
Firm
Abbott Diabetes Care, Inc.
FEI Number
3002803444
Product Code
QLG
Status
Open, Classified
Root Cause
Process control
Initiated
November 24, 2025
Posted
January 14, 2026
Address
1360 S Loop Rd, Alameda, CA, 94502-7000

Description

FreeStyle Libre 3 Plus Sensor REF 72080-01 UDI-DI code: 00357599819002 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.

Reason

Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results

Action

On 11/24/2025, the firm communicated "Urgent Medical Device Correction" letter to customers and Healthcare Providers that Abbott has recently identified that certain FreeStyle Libre 3 and FreeStyle Libre 3 Plus sensors provide incorrect low glucose readings. We learned about this through our ongoing product monitoring, where Libre 3 and Libre 3 Plus sensor users reported situations where they received incorrect low glucose readings. Customers are instructed: -If you are currently wearing or have a FreeStyle Libre 3 / FreeStyle Libre 3 Plus sensor that has been confirmed as potentially affected on www.FreeStyleCheck.com or by a customer service representative, immediately discontinue use and dispose of the affected sensor(s). You can request a replacement for any potentially affected sensor(s) on www.FreeStyleCheck.com. Select CONFIRM SENSOR SERIAL NUMBER and enter a valid serial number. If your sensor is potentially impacted, you will be instructed to enter your contact information so a replacement product can be sent to you at no cost. -Use a blood glucose meter or the built-in meter in your FreeStyle Libre 3 Reader to make treatment decisions when your sensor readings don t match your symptoms or expectations. -If you are wearing a FreeStyle Libre 3 and FreeStyle Libre 3 Plus sensor, you can find the serial number in the app or reader. The serial number can also be found on the label on the bottom of the sensor applicator or carton. (If you are using a sensor with a connected insulin delivery device, please refer to the connected insulin delivery device user manual on how to locate the sensor serial number.) For questions/to report any adverse reactions/quality problems experienced, call Abbott Customer Service at 1-833-815-4273, available seven days a week from 8 a.m. to 8 p.m. Eastern Time. Agents are available 24/7 through live chat at https://www.freestyle.abbott/us-en/support/contact-us.html.

Distribution

U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.

Quantity

8,062 sensors