FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Recall
- Recall Number
- Z-1020-2026
- Event Number
- 98049
- Firm
- Abbott Diabetes Care, Inc.
- FEI Number
- 3002803444
- Product Code
- QLG
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- November 24, 2025
- Posted
- January 14, 2026
- Address
- 1360 S Loop Rd, Alameda, CA, 94502-7000
Description
FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
On 11/24/2025, the firm communicated "Urgent Medical Device Correction" letter to customers and Healthcare Providers that Abbott has recently identified that certain FreeStyle Libre 3 and FreeStyle Libre 3 Plus sensors provide incorrect low glucose readings. We learned about this through our ongoing product monitoring, where Libre 3 and Libre 3 Plus sensor users reported situations where they received incorrect low glucose readings. Customers are instructed: -If you are currently wearing or have a FreeStyle Libre 3 / FreeStyle Libre 3 Plus sensor that has been confirmed as potentially affected on www.FreeStyleCheck.com or by a customer service representative, immediately discontinue use and dispose of the affected sensor(s). You can request a replacement for any potentially affected sensor(s) on www.FreeStyleCheck.com. Select CONFIRM SENSOR SERIAL NUMBER and enter a valid serial number. If your sensor is potentially impacted, you will be instructed to enter your contact information so a replacement product can be sent to you at no cost. -Use a blood glucose meter or the built-in meter in your FreeStyle Libre 3 Reader to make treatment decisions when your sensor readings don t match your symptoms or expectations. -If you are wearing a FreeStyle Libre 3 and FreeStyle Libre 3 Plus sensor, you can find the serial number in the app or reader. The serial number can also be found on the label on the bottom of the sensor applicator or carton. (If you are using a sensor with a connected insulin delivery device, please refer to the connected insulin delivery device user manual on how to locate the sensor serial number.) For questions/to report any adverse reactions/quality problems experienced, call Abbott Customer Service at 1-833-815-4273, available seven days a week from 8 a.m. to 8 p.m. Eastern Time. Agents are available 24/7 through live chat at https://www.freestyle.abbott/us-en/support/contact-us.html.
U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.
1,060,723 sensors