FDA Recall Open, Classified

Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Recall: Z-1016-2023 · Initiated December 6, 2022

Recall

Recall Number
Z-1016-2023
Event Number
91344
Firm
Future Diagnostics Solutions B.V. Nieuweweg 279 Wijchen Netherlands
FEI Number
3001857593
Product Code
CEW
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
December 6, 2022
Posted
January 25, 2023

Description

Future Diagnostics STAT-IntraOperative-Intact-PTH (STAT-IO-I-PTH) Immunoassay kit REF 4K-IPT-00. in vitro diagnostic.

Reason

Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.

Action

On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to [email protected].

Distribution

Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom

Quantity

342 kits