FDA Recall Terminated

Infant Warmer System (IWS)

Recall: Z-1015-2017 · Initiated August 4, 2016

Recall

Recall Number
Z-1015-2017
Event Number
75210
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
FMT
Status
Terminated
Root Cause
Use error
Initiated
August 4, 2016
Terminated
July 12, 2018
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Infant Warmer System (IWS)

Reason

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Action

This recall strategy includes the depth of the recall, plan for public warning and effectiveness check. The strategy outlined below addresses these elements for this particular recall. The plan for public warning is to contact all customers with potentially affected product by means of an Urgent Medical Device Correction letter. The letter will instruct customers that they can continue to use the device by following instructions on the Addendum to the Service Manual. The Service Manual Addendum provides instructions for checking and tightening the screws. In the US the letter will be sent by a traceable means [either US Postal Service with receipt acknowledgement (registered mail) or equivalent courier service such as FEDEX]. This letter will be sent to: Nurse Manager, Labor & Delivery/NICU; Manager, Bio-Medical Engineering; and Director of Risk Management.

Distribution

This product has been distributed domestically and internationally .

Quantity

32,073 (US = 24,361; OUS = 7,712)