FDA Recall Terminated

Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated protocol converter used to concert the serial data from a DINAMAP Monitor to that of the APEXPRO telemetry transmitter.

Recall: Z-1015-04 · Initiated January 27, 2004

Recall

Recall Number
Z-1015-04
Event Number
28997
Firm
GE Medical Systems Information Technologies
FEI Number
1030184
Product Code
MHX
Status
Terminated
Root Cause
Other
Initiated
January 27, 2004
Posted
July 20, 2004
Terminated
October 5, 2009
Address
4502 Woodland Corporate Blvd, Tampa, FL, 33614-2422

Description

Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated protocol converter used to concert the serial data from a DINAMAP Monitor to that of the APEXPRO telemetry transmitter.

Reason

The potential exists for external electrical noise to be processed as irregular pacing thereby masking potential asystole events.

Action

The firm mailed an 'Urgent Medical Device Correction' letter to consignees on January 27, 2004 advising them of the external electrical noise problem and informing them to not use the Dinalink ILC1931 with patients with pace makers until a fix or replacement can be provided. By 5/17/2004 corrective replacement cables had been provided to all consignees.

Distribution

Product was distributed to 7 domestic hospitals in PA, SC, GA, FL, WI, SD AND WY.

Quantity

36