Drager Infinity CentralStation Software versions VG2.1.2 and VG2.1.2 SU12. For patient monitoring with Infinity M300 wireless telemetry devices.

Recall: Z-1012-2022 · Initiated March 4, 2022

Recall

Recall Number
Z-1012-2022
Event Number
89877
Firm
Draegar Medical Systems, Inc.
FEI Number
3005783425
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
March 4, 2022
Address
6 Tech Dr, Andover, MA, 01810-2434

Description

Drager Infinity CentralStation Software versions VG2.1.2 and VG2.1.2 SU12. For patient monitoring with Infinity M300 wireless telemetry devices.

Reason

Software issue resulting in temporary loss of central monitoring functionality.

Action

A notification signed 03/08/2022, issued via certified mail on 03/11/2022 and 03/24/2022, notified customers of the impacted software versions, that a software update will be deployed to correct the issue, and provided interim instructions regarding use of USB devices. Customers are to notify all users and return the acknowledgement form. For questions please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349).

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IA, ID, IL, IN, LA, MN, MS, NC, NE, NH, NY, OH, PA, SD, TN, TX, WA, WI, and 58 additional foreign countries.

Quantity

1950 (1574 OUS + 376 US)