Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.
Recall
- Recall Number
- Z-1012-2015
- Event Number
- 70086
- Firm
- Aesculap, Inc.
- FEI Number
- 2916714
- Product Code
- JXG
- Status
- Terminated
- Root Cause
- Labeling False and Misleading
- Initiated
- December 17, 2014
- Posted
- January 26, 2015
- Terminated
- December 13, 2016
- Address
- 3773 Corporate Pkwy, Center Valley, PA, 18034-8217
Description
Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.
Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
Aesculap sent an Important Correction and Removal and Field Safety Notification letter, dated December 17, 2014, to all affected customers. The letter identiifed the affected product, problem, and actions to be taken. Customers were requested to remove from inventory and contact Aesculap regarding return of the affected devices and acknowledgement form.
Worldwide Distribution - US Nationwide
772