FDA Recall Open, Classified

Sensi Disc Amoxicillin with Clavulanic Acid 20/10 g, Catalog No. 231628 and 231629. These discs are used for semi-quantitative in vitro susceptibility testing

Recall: Z-1011-2024 · Initiated January 8, 2024

Recall

Recall Number
Z-1011-2024
Event Number
93789
Firm
Becton Dickinson & Co.
FEI Number
1119779
Product Code
JTN
Status
Open, Classified
Root Cause
Process control
Initiated
January 8, 2024
Posted
February 8, 2024
Address
7 Loveton Cir, Sparks, MD, 21152-9212

Description

Sensi Disc Amoxicillin with Clavulanic Acid 20/10 g, Catalog No. 231628 and 231629. These discs are used for semi-quantitative in vitro susceptibility testing

Reason

There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Action

On January 8, 2024, the firm notified customers through URGENT Medical Device Product Correction letters. Users were instructed to cease use of catalog number 291270, and to refrain from conducting AST for H. influenzae with impacted Sensi-Discs. Customers should destroy any units of catalog number 291270 in their inventory. Customers should share and post the recall notification within their facility network and forward to any customers to whom they may have distributed the product. Future lots of affected Sensi-Disc products (excluding catalog number 291270) will be labeled with a sticker notifying customers "This product should not be used for the semi-quantitative in vitro susceptibility testing of Haemophilus influenzae." Identify impacted lot numbers: https://legacy.bd.com/alerts-notices/IDS-23-4851.asp For customer questions: Phone: 1-844-823-5433 Say "Recall when prompted Mon-Fri 8:00am and 5:00pm CT or Email: [email protected]

Distribution

Domestic distribution nationwide. International distribution worldwide.

Quantity

2,363,168 total units