FDA Recall Terminated

INTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG Some items starting with codes 0022530

Recall: Z-1010-2014 · Initiated October 11, 2013

Recall

Recall Number
Z-1010-2014
Event Number
66500
Firm
Zimmer, Inc.
FEI Number
1000220733
Product Code
NDJ
Status
Terminated
Root Cause
Packaging
Initiated
October 11, 2013
Posted
February 20, 2014
Terminated
April 14, 2017
Address
345 E Main St, Warsaw, IN, 46580-2746

Description

INTERLOCKING IM SCREW 4.5 and INTERLOCKING IM SCREW LG Some items starting with codes 0022530

Reason

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Action

Distributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.

Quantity

7,044,680 total