FDA Recall Terminated

RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.

Recall: Z-1008-2022 · Initiated March 9, 2022

Recall

Recall Number
Z-1008-2022
Event Number
89898
Firm
Roche Diabetes Care, Inc.
FEI Number
3011393376
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
March 9, 2022
Terminated
January 22, 2025
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.

Reason

Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.

Action

Urgent Medical Device Correction Notice 22-001 was issued March 09, 2022 via email. The software issue was resolved with the cloud-based software update to version 2.5.5 on February 24, 2022. No further action required. Customers to submit notification acknowledgement. Contact your Roche Account Manager or contact Accu-Chek Customer Care at 1-800-628-3346 with questions.

Distribution

US Nationwide Distribution: AK, AZ, CA, CO, DC, FL, GA, IL, IN, MN, NC, NJ, NY, PA, RI, SC, TX, WI

Quantity

81 registered accounts