FDA Recall Terminated

MAQUET Servo Humidifier 163; Model No.: 01-06-8125-8, catalog no. XKC01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.

Recall: Z-1008-2015 · Initiated December 4, 2014

Recall

Recall Number
Z-1008-2015
Event Number
70127
Firm
Teleflex, Inc.
FEI Number
3011369183
Product Code
BYD
Status
Terminated
Root Cause
Process control
Initiated
December 4, 2014
Posted
February 6, 2015
Terminated
March 21, 2017
Address
550 E Swedesford Rd, Ste 400, Wayne, PA, 19087-1601

Description

MAQUET Servo Humidifier 163; Model No.: 01-06-8125-8, catalog no. XKC01-06-8125-8. A heat and moisture condenser (artificial nose) is a device intended to be positioned over a tracheotomy (a surgically created opening in the throat) or tracheal tube (a tube inserted into the trachea) to warm and humidify gases breathed in by a patient.

Reason

Connector mount cracks which may lead to a leak failure during use and necessitate immediate replacement in the breathing circuit. If leak exists and is left untreated, patient may be deprived of adequate anesthetic gases over a period of time and serious adverse health consequences may occur.

Action

An Urgent - Field Safety Notice, dated December 4, 2014, was sent to end users that identified the product, problem, and action to be taken. Customers were requested to immediately check inventory, discontinue use and distribution, quarantine. Customers were to return the Acknowledgement form and request a return authorization number from their local Maquet representative for affected devices. A credit will be issued upon receipt of the returned affected product. For questions or for support concerning this issue, a local Maquet representative should be contacted at [email protected].

Distribution

Distributed in the states of FL, MI, WV, and KS.

Quantity

18