FDA Recall Completed

Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122

Recall: Z-1007-2026 · Initiated July 14, 2022

Recall

Recall Number
Z-1007-2026
Event Number
98102
Firm
Philips Ultrasound, LLC
FEI Number
3019216
Product Code
IYN
Status
Completed
Root Cause
Software Design Change
Initiated
July 14, 2022
Posted
December 31, 2025
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122

Reason

Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.

Action

On July 18, 2022, Phillips issued a product notification to affected consignees. Philips ask consignees to take the following actions: 1. If the X5-1c transducer is connected to the right-most port (when viewing from the front), this problem will not occur. Please ensure your X5-1c is connected to the right-most connector until the software update (FCO79500561) has been installed. 2. Please retain this letter with your product documentation until the software update (FCO79500561) has been implemented on your system. 3. Please circulate this notice to all users of this device so that they are aware of the issue. 4. A Philips Field Service Engineer will contact you to schedule the installation of the software update to resolve this issue (reference FCO79500561).

Distribution

Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Argentina, Australia, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of Latvia, Lithuania, Morocco, Netherlands , New Zealand, Oman, Panama, Philippines, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.

Quantity

1,721 systems