FDA Recall Terminated

Vitatron T-Series Dual Chamber pacemaker T60 DR, model T60A1. The Vitatron C60 DR is a dual chamber rate responsive pacemaker (activity sensing using an accelerometer) for permanent atrial and ventricular pacing. T-Series software VSF07/VSF08 Version 1.2. Vitatron is a division of Medtronic, Inc., Sales Office is Vitatron USA, 7000 Central Ave., NE, Minneapolis, MN 55423-3576. The head office is Vitatron B.V., P.O. Box 5227, 6802 EE Arnhem, The Netherlands

Recall: Z-1000-2007 · Initiated February 5, 2007

Recall

Recall Number
Z-1000-2007
Event Number
37587
Firm
Medtronic Inc. Cardiac Rhythm Managment
FEI Number
2182208
Product Code
NVZ
Status
Terminated
Root Cause
Software design
Initiated
February 5, 2007
Posted
July 7, 2007
Terminated
October 16, 2008
Address
7000 Central Ave Ne, Minneapolis, MN, 55432-3568

Description

Vitatron T-Series Dual Chamber pacemaker T60 DR, model T60A1. The Vitatron C60 DR is a dual chamber rate responsive pacemaker (activity sensing using an accelerometer) for permanent atrial and ventricular pacing. T-Series software VSF07/VSF08 Version 1.2. Vitatron is a division of Medtronic, Inc., Sales Office is Vitatron USA, 7000 Central Ave., NE, Minneapolis, MN 55423-3576. The head office is Vitatron B.V., P.O. Box 5227, 6802 EE Arnhem, The Netherlands

Reason

Pacing rate: Software anomaly affecting Vitatron dual chamber C-series and T-series pacemakers, related to automatic retrograde conduction testing, can inhibit the device from pacing if the patient's intrinsic rate falls below the programmed lower rate.

Action

Advisory letters to physicians were mailed in February 2007. Sales respresentatives will follow-up with physicians by visit. A programmer software update will be released upon regulatory approval. The Advisory Letter provided short-term direction to reduce problems pending software upgrades.

Distribution

within the US to include: AL, AR, AZ, CA, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, NC, ND, NJ, NY, OH, OK, SC, TX.

Quantity

6,259