Vitatron T-Series Dual Chamber pacemaker T60 DR, model T60A1. The Vitatron C60 DR is a dual chamber rate responsive pacemaker (activity sensing using an accelerometer) for permanent atrial and ventricular pacing. T-Series software VSF07/VSF08 Version 1.2. Vitatron is a division of Medtronic, Inc., Sales Office is Vitatron USA, 7000 Central Ave., NE, Minneapolis, MN 55423-3576. The head office is Vitatron B.V., P.O. Box 5227, 6802 EE Arnhem, The Netherlands
Recall
- Recall Number
- Z-1000-2007
- Event Number
- 37587
- Firm
- Medtronic Inc. Cardiac Rhythm Managment
- FEI Number
- 2182208
- Product Code
- NVZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- February 5, 2007
- Posted
- July 7, 2007
- Terminated
- October 16, 2008
- Address
- 7000 Central Ave Ne, Minneapolis, MN, 55432-3568
Description
Vitatron T-Series Dual Chamber pacemaker T60 DR, model T60A1. The Vitatron C60 DR is a dual chamber rate responsive pacemaker (activity sensing using an accelerometer) for permanent atrial and ventricular pacing. T-Series software VSF07/VSF08 Version 1.2. Vitatron is a division of Medtronic, Inc., Sales Office is Vitatron USA, 7000 Central Ave., NE, Minneapolis, MN 55423-3576. The head office is Vitatron B.V., P.O. Box 5227, 6802 EE Arnhem, The Netherlands
Pacing rate: Software anomaly affecting Vitatron dual chamber C-series and T-series pacemakers, related to automatic retrograde conduction testing, can inhibit the device from pacing if the patient's intrinsic rate falls below the programmed lower rate.
Advisory letters to physicians were mailed in February 2007. Sales respresentatives will follow-up with physicians by visit. A programmer software update will be released upon regulatory approval. The Advisory Letter provided short-term direction to reduce problems pending software upgrades.
within the US to include: AL, AR, AZ, CA, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, NC, ND, NJ, NY, OH, OK, SC, TX.
6,259