Zoll M Series External Defibrillator
Recall
- Recall Number
- Z-1000-03
- Event Number
- 26590
- Firm
- Zoll Medical Corporation
- FEI Number
- 1220908
- Product Code
- MKJ
- Status
- Terminated
- Root Cause
- Other
- Initiated
- June 23, 2003
- Posted
- July 10, 2003
- Terminated
- December 29, 2020
- Address
- 32 Second Avenue, Burlington, MA, 01803
Description
Zoll M Series External Defibrillator
Defibrillator may fail to charge or hold its charge above energies of 75 joules
Zoll Medical issued a letter on 6/25/03 to domestic consignees and 6/26/03 to international consignees on 6/26/03. Letters were issued Certified Mail Return Receipt requested. Instructions are provided for Biomedical to charge device and then dump the charge. A damaged device will fail to charge and issue an error message. Users then are requested to return device. Firm will initiate telephone contact on 6/30/03 to all accounts to verify receipt of notice and confirmation of testing per the instructions.
Nationwide Foreign: Emirates, Australia, Brussels, Brazil, Canada, China, Columbia, Chezche, Germany, Spain, Finland, France, Greece, Hong Kong, India, Indonesia, Italy, Japan, Korea, Kuwait, Latvia, Mexico, Netherlands, Poland,Romania, Saudi Arabia, Syrai, Thailand, Turkey,
8,142 units