FDA Recall Terminated

Powered stretcher chair used during facial and eye surgeries. One chair per package.

Recall: Z-0996-2015 · Initiated November 14, 2014

Recall

Recall Number
Z-0996-2015
Event Number
69671
Firm
TransMotion Medical Inc
FEI Number
1027229
Product Code
GBB
Status
Terminated
Root Cause
Component design/selection
Initiated
November 14, 2014
Posted
January 16, 2015
Terminated
July 11, 2015
Address
1441 Wolf Creek Trail, Sharon Center, OH, 44274-0302

Description

Powered stretcher chair used during facial and eye surgeries. One chair per package.

Reason

The firm is finding that the stretcher chair controllers are moving the chairs involuntarily.

Action

TransMotion Medical Inc. sent an Urgent Medical Device Recall letter dated November 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: Discontinue use of the device and contact TransMotion Medical Inc. for a replacement. Suspected faulty pendants must be destroyed by severing the cord and discarding the control. Replacement of the pendant is a long-term solution to the issue due to replacement pendants being not of the same type or design as the failed pendant. Accompanying the recall letter will be an acknowledgment and receipt form that is required to be completed by the customer and returned to TMM via fax or email. This form will confirm that the product resides at the customer site where the unit was originally shipped. If the unit is no longer present at the customer site the customer can indicate this on the form and provide further information as to the whereabouts of the device if known. Ship replacement pendant control to the customer free of charge with installation instructions as to how to replace the pendant on the device. The firm will track non-responders and initiate follow-up contact. For further questions please call Phone: 330-239-4192 ext. 175 or Toll free: 1-866-860-8447 (7:30 AM to 4:00 PM Eastern Time Monday through Friday)

Distribution

Worldwide Distribution - US Distribution and the countries including Canada, Australia, Chile, China, Costa Rica, Germany, Algeria, Israel, Italy, South Korea, Malaysia, Nigeria, Netherlands, Poland, Portugal, Saudi Arabia, Singapore, Thailand, and Venezuela.

Quantity

1,127 units