FDA Recall Open, Classified

Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.

Recall: Z-0995-2026 · Initiated November 26, 2025

Recall

Recall Number
Z-0995-2026
Event Number
98027
Firm
Cook Incorporated
FEI Number
1820334
Product Code
DYB
Status
Open, Classified
Root Cause
Process control
Initiated
November 26, 2025
Posted
December 23, 2025
Address
750 N Daniels Way, Bloomington, IN, 47404-9120

Description

Rsch-Uchida Transjugular Liver Access Sets, Reference Part Numbers: RUPS-100; The Rsch-Uchida Transjugular Liver Access Set is intended for transjugular liver access in diagnostic and interventional procedures.

Reason

Cook Medical identified that products supplied in the affected device lots may have been manufactured out of specification. Impacted products may have been cut to the incorrect length, not correctly trimmed, and not inspected properly.

Action

Cook Medical issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 11/26/2025 by email and/or letter delivered via a courier. The notice explained the issue, risk to health, and requested the identification and return of the affected product. The notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the affected devices have been transferred. For questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.

Distribution

Worldwide distribution.

Quantity

23 units