Cyberonics VNS Therapy Demipulse Duo Generator, Model 104; an implantable, multiprogrammable, vagus nerve stimulator.

Recall: Z-0995-2009 · Initiated February 7, 2008

Recall

Recall Number
Z-0995-2009
Event Number
47442
Firm
Cyberonics, Inc
FEI Number
1000117004
Product Code
LYJ
Status
Terminated
Root Cause
Software design
Initiated
February 7, 2008
Posted
April 22, 2009
Terminated
February 28, 2011
Address
100 Cyberonics Blvd, Houston, TX, 77058-2069

Description

Cyberonics VNS Therapy Demipulse Duo Generator, Model 104; an implantable, multiprogrammable, vagus nerve stimulator.

Reason

Reset/disabling of the VNS Therapy Demipulse Generator and Demipulse Duo Generator due to magnet interference, resulting in the loss of stimulation.

Action

Cyberonics notified consignees of recall via letter on 02/07/08. Consignees were notified of reset issue in which during a strong magnetic field, the software may report an error even though no error occurred. This error causes the generator to reset, which disables the generator (i.e., no stimulation delivered). Physicians should inform any potential and currently implanted patients of the possibility of this event. Patients should continue with their regularly scheduled visits as determined by their physician. If this event is suspected, physicians should interrogate the device as described in the Physicians Manual to confirm the generator has been reset and Cyberonics should be contacted for any additional instruction.

Distribution

Worldwide Distribution - USA, Belgium, Germany, Norway, Spain, Sweden, the Netherlands, and United Kingdom.

Quantity

108 units.