FDA Recall Terminated

Integra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Recall: Z-0994-2014 · Initiated December 5, 2013

Recall

Recall Number
Z-0994-2014
Event Number
67406
Firm
Integra LifeSciences Corp.
FEI Number
3003418325
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
December 5, 2013
Posted
February 20, 2014
Terminated
January 26, 2017
Address
311 Enterprise Dr, Plainsboro, NJ, 08536-3344

Description

Integra Cusa Excel Rx Only CEM" Nosecone for CUSA Excel System Product Usage: Intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator The CUSA Electrosurgical Module (CEM) provides Desiccate Coagulation waveform electrosurgical capability to CUSA Excel handpieces. The CUSA handpiece with the CEM Nosecone works in conjunction with the Force FX generator in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.

Reason

The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.

Action

Integra sent an Urgent: Medical Device Correction Letter/Acknowledgement Return Form dated December 5, 2013 via traceable courier to their customers. The notification: advises customers of the nature of the issue and how it could affect CEM Nosecone use, advises customers what to do if the issue occurs and requests customers complete and return the Acknowledgment Return Form. The form should be faxed to [email protected] or 1-609-275-9445. Any questions can be directed to customer service at 855-532-1723.

Distribution

Worldwide Distribution - USA Nationwide

Quantity

17253 boxes