IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836
Recall: Z-0994-05
·
Initiated April 22, 2005
Recall
- Recall Number
- Z-0994-05
- Event Number
- 32010
- Firm
- DPC Cirrus
- FEI Number
- 1000122204
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 22, 2005
- Posted
- July 16, 2005
- Terminated
- June 29, 2006
- Address
- 62 Flanders Bartley Rd, Flanders, NJ, 07836-4715
Description
IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836
Reason
The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.
Action
The firm became aware of the problem on 4/19 and conducted tests on 4/22/05. A service alert for field service engineers and Immulite 2000 distributors was prepared on 4/22/05.
Distribution
worldwide
Quantity
There are 941 units in the United States and 3014 units in the international market.