IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836

Recall: Z-0994-05 · Initiated April 22, 2005

Recall

Recall Number
Z-0994-05
Event Number
32010
Firm
DPC Cirrus
FEI Number
1000122204
Product Code
JJE
Status
Terminated
Root Cause
Other
Initiated
April 22, 2005
Posted
July 16, 2005
Terminated
June 29, 2006
Address
62 Flanders Bartley Rd, Flanders, NJ, 07836-4715

Description

IMMULITE 2000. An Automated, Quantitative Immunoassay Analyzer for In Vitro Diagnostic Use. Manufactured by DPC Cirrus Inc., Flanders, NJ 07836

Reason

The Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.

Action

The firm became aware of the problem on 4/19 and conducted tests on 4/22/05. A service alert for field service engineers and Immulite 2000 distributors was prepared on 4/22/05.

Distribution

worldwide

Quantity

There are 941 units in the United States and 3014 units in the international market.