FDA Recall Open, Classified

PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701

Recall: Z-0993-2022 · Initiated March 15, 2022

Recall

Recall Number
Z-0993-2022
Event Number
89823
Firm
Baxter Healthcare Corporation
FEI Number
1417572
Product Code
KDI
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
March 15, 2022
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701

Reason

This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be returned to the original value immediately upon selecting the Same Patient function. The Gain/Loss Limit value will be returned to the original value when the patient weight and/or hematocrit is changed during treatment.

Action

The firm notified its consignees be letter on 03/15/2022. The following instructions were provided: "1. Operators may continue to safely use the PrisMax System. To ensure patient safety, there are two approaches to address the anomalies: Approach 1: Choose new patient at every filter set change. This will require re-input of all prescription settings and therapy will begin without prescription settings and treatment data from the previous filter set. Approach 2: Choose Same Patient when changing a filter set. Check and correct the Gain/Loss Limit or RDL prior to starting treatment. See Attachment A to change the Gain/Loss Limit See Attachment B to change the RDL The device should only be used by a trained operator per the instructions in the Operator s Manual. Please ensure that every operator of this device is made aware of this Urgent Medical Device Correction. 2. When the software upgrade becomes available, a local Baxter representative will contact your facility to determine the correction plan and schedule the upgrade for impacted devices. Your facility will be receiving this upgrade from Baxter at no charge. 3. Acknowledge the receipt of this notification by completing a reply form on the customer portal."

Distribution

Worldwide distribution - US Nationwide and the countries of Mexico, Brazil, Jamaica, Trinidad and Tobago, Canada, Australia, Singapore, Hong Kong, Vietnam, New Zealand, Korea, Austria, Belgium, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Quantity

3513 units