ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan
Recall
- Recall Number
- Z-0992-04
- Event Number
- 29044
- Firm
- R & D Systems, Inc
- FEI Number
- 2182501
- Product Code
- JPK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 15, 2004
- Posted
- July 20, 2004
- Terminated
- June 24, 2004
- Address
- 614 Mckinley Pl Ne, Minneapolis, MN, 55413-2610
Description
ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan
Clots in hematology controls resulted in low red blood cell counts when used in blood analyzers because control values were outside the expected range due to the clotting which took place after manufacturing.
Recall letters dated April 15, 2004 instructed consignees not to use the products to verify their analyzers and stated that the recalling firm is working on providing replacement products.
The products were shipped to California, France and Japan.
29494 vials of the normal controls and 15196 of the high controls