FDA Recall Terminated

ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan

Recall: Z-0992-04 · Initiated April 15, 2004

Recall

Recall Number
Z-0992-04
Event Number
29044
Firm
R & D Systems, Inc
FEI Number
2182501
Product Code
JPK
Status
Terminated
Root Cause
Other
Initiated
April 15, 2004
Posted
July 20, 2004
Terminated
June 24, 2004
Address
614 Mckinley Pl Ne, Minneapolis, MN, 55413-2610

Description

ABX Diagnostics Minotrol 16 Hematology Controls, Normal and High, in vials distributed by ABX Diagnostics, Irvine, CA; and Horiba LC-TROL 16 (Hematology Reference Controls), Normal and High, in vials distributed by Horiba Ltd., Kyoto, Japan

Reason

Clots in hematology controls resulted in low red blood cell counts when used in blood analyzers because control values were outside the expected range due to the clotting which took place after manufacturing.

Action

Recall letters dated April 15, 2004 instructed consignees not to use the products to verify their analyzers and stated that the recalling firm is working on providing replacement products.

Distribution

The products were shipped to California, France and Japan.

Quantity

29494 vials of the normal controls and 15196 of the high controls