FDA Recall Terminated

VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc7700 8-port, Cat. #W0447; (4) dc7700 8-port (refurbished), Cat. #W0447R; (5) dc7800, Cat. #W0449; (6) dc7800 (refurbished), Cat. #W0449R (7) rp5700, Cat. #W0452; (8) rp5700 (refurbished), Cat. #W0452R; (9) rp5800 XPE, Cat. 413642; (10) rp5800WES7, Cat. #413862; (11) d530 6-port, Cat. #W0438; (12) d530 6-port (refurbished), Cat. #W0438R; and (13) d530 4-port, Cat. #W0437. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Recall: Z-0991-2013 · Initiated February 18, 2013

Recall

Recall Number
Z-0991-2013
Event Number
64394
Firm
Biomerieux Inc
FEI Number
1950204
Product Code
LTW
Status
Terminated
Root Cause
Software design
Initiated
February 18, 2013
Posted
March 25, 2013
Terminated
August 14, 2014
Address
595 Anglum Rd, Hazelwood, MO, 63042-2320

Description

VITEK 2 Compact computer system when equipped with the following Hewlett-Packard PC models: (1) dc7100 8-port, Cat. #W0441; (2) dc7100 8-port (refurbished), Cat. #W0441R; (3) dc7700 8-port, Cat. #W0447; (4) dc7700 8-port (refurbished), Cat. #W0447R; (5) dc7800, Cat. #W0449; (6) dc7800 (refurbished), Cat. #W0449R (7) rp5700, Cat. #W0452; (8) rp5700 (refurbished), Cat. #W0452R; (9) rp5800 XPE, Cat. 413642; (10) rp5800WES7, Cat. #413862; (11) d530 6-port, Cat. #W0438; (12) d530 6-port (refurbished), Cat. #W0438R; and (13) d530 4-port, Cat. #W0437. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Reason

There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.

Action

bioMerieux sent an "URGENT PRODUCT CORRECTION NOTICE" dated February 20, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Emails were issued by the recalling firm on 2/18/2013 and 2/19/2013 to their subsidiaries and to distributors. Letters began issuing to customers via e-mail on 2/20/2013. Contact bioMerieux Clinical Customer Service at (800) 682-2666 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaidjian, Bahrain, Belarus, Belgium, Benin, Bolivia, Bosnia, Brazil, Bulgaria, Burkina Faso, Camaroun, Canada, Chad, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Democratic Republic of Congo, Demark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Guinea, Herzegovina, Honduras, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mali, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Netherlands, Netherlands, Nicaragua, Niger Niamey, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, and Yemen.

Quantity

For all 3 computers types: 16,736 originally distributed; 554 refurbished and redistributed