FDA Recall Open, Classified

Flow-i C40 Anesthesia System, model 6677400

Recall: Z-0989-2024 · Initiated December 11, 2023

Recall

Recall Number
Z-0989-2024
Event Number
93736
Firm
Getinge Usa Sales Inc
FEI Number
3013876692
Product Code
BSZ
Status
Open, Classified
Root Cause
Software design
Initiated
December 11, 2023
Posted
February 6, 2024
Address
1 Geoffrey Way, Wayne, NJ, 07470-2035

Description

Flow-i C40 Anesthesia System, model 6677400

Reason

Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.

Action

An Urgent Medical Device Correction notice dated December 21, 2023 was issued via FedEx. The letter advises customers to examine inventory immediately to determine if you have any affected product. Directions are outlined for 1) Users with system version 4.8 and above with a Flow Anesthesia System connected to Connected Services directly via the Ethernet port on the Flow Anesthesia System (i.e., without a Getinge Connect module) and for 2) Users with system version 4.7 and below with Flow Anesthesia System connected to Remote Services via ethernet cable. Customers are asked to complete and return the MEDICAL DEVICE - CORRECTION RESPONSE FORM. Information should be forwarded to all current and potential system users with in your hospital / facility. Distributors are to forward the notification to customers for appropriate action. Contact your Getinge representative or call Getinge Customer Service at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) with questions.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.

Quantity

16 (US)