Alden Classic 55 Toric Daily Wear Soft contact; Catalog Number: AOCLT55 Product Usage: The ALDEN TORIC LENS for daily wear is indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with refractive ametropia (myopia or hyperopia) and/or possesses refractive astigmatism not exceeding 10 Diopters.
Recall
- Recall Number
- Z-0989-2018
- Event Number
- 79377
- Firm
- Alden Optical
- FEI Number
- 1315522
- Product Code
- LPL
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- November 3, 2017
- Terminated
- June 19, 2018
- Address
- 6 Lancaster Pkwy, Lancaster, NY, 14086-9713
Description
Alden Classic 55 Toric Daily Wear Soft contact; Catalog Number: AOCLT55 Product Usage: The ALDEN TORIC LENS for daily wear is indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with refractive ametropia (myopia or hyperopia) and/or possesses refractive astigmatism not exceeding 10 Diopters.
Contact lenses lack sterility assurance.
Consignees were initially called and told over the phone not to further distribute the contact lenses. For two consignees whose patients could not be reached by phone, a letter was sent. The consignee letter states the specific product and lot should not be further distributed due to the lack of sterility assurance. A business reply card was provided with the letter and contacts should be returned to the manufacturer using a label provided by the firm.
Worldwide Distribution - US Nationwide in the states:of CA, FL, GA, ID, IL, IN, KS, KY, MD, MI, MO, MS, NC, NJ, NM, NV, NY, OH, OR, PA, TX, VA, WI, WY and the country of Canada.
1