FDA Recall Terminated

i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448

Recall: Z-0989-2017 · Initiated November 23, 2016

Recall

Recall Number
Z-0989-2017
Event Number
76102
Firm
HEALTH ROBOTICS S.R.L. Via Sebastian Altmann 9 A 39100 Bolzano Italy
FEI Number
3006153342
Product Code
NEP
Status
Terminated
Root Cause
Software design
Initiated
November 23, 2016
Posted
January 10, 2017
Terminated
March 15, 2017

Description

i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448

Reason

The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.

Action

Omincell mailed an Urgent Medical Device Field Correction Notification letter to affected customers on 11/21/2016. The letter provided information of the event and instructions on how to continue operating the device to prevent risks. The issue will be corrected with a software update.

Distribution

Shipped throughout United States.

Quantity

121 systems