FDA Recall
Terminated
i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448
Recall: Z-0989-2017
·
Initiated November 23, 2016
Recall
- Recall Number
- Z-0989-2017
- Event Number
- 76102
- Firm
- HEALTH ROBOTICS S.R.L. Via Sebastian Altmann 9 A 39100 Bolzano Italy
- FEI Number
- 3006153342
- Product Code
- NEP
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 23, 2016
- Posted
- January 10, 2017
- Terminated
- March 15, 2017
Description
i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System Device Listing Number E454448
Reason
The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.
Action
Omincell mailed an Urgent Medical Device Field Correction Notification letter to affected customers on 11/21/2016. The letter provided information of the event and instructions on how to continue operating the device to prevent risks. The issue will be corrected with a software update.
Distribution
Shipped throughout United States.
Quantity
121 systems