FDA Recall Terminated

Deltec Cozmo Insulin Pump, Model 1700 containing 1800 software. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients, but can also be used in a healthcare facility.

Recall: Z-0989-2009 · Initiated November 20, 2008

Recall

Recall Number
Z-0989-2009
Event Number
50872
Firm
Smiths Medical MD, Inc.
FEI Number
2183502
Product Code
LZG
Status
Terminated
Root Cause
Software design
Initiated
November 20, 2008
Posted
April 16, 2009
Terminated
June 5, 2012
Address
1265 Grey Fox Rd, Saint Paul, MN, 55112-6929

Description

Deltec Cozmo Insulin Pump, Model 1700 containing 1800 software. Deltec Cozmo Insulin Pumps are ambulatory electromechanical pumps used for delivering insulin to patients with diabetes. The pump is used mainly by homecare patients, but can also be used in a healthcare facility.

Reason

Smiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered was not accurately displayed by the Pump. The amount of extended Bolus displayed on the Pump Home Screen 2 and in the Bolus Summary Report is less than what w

Action

Consignees were sent a Smiths Medical "Urgent Device Recall" letter dated November 20, 2008. The letter was addressed to Distributors, Clinicians, Patients and other Customers. The letter included a Description of the Problem, Advice on Action to be Taken by the User and requested the return of the Confirmation Form. In particular, users were given the option to check the "Replacement Pump" option on the Confirmation Form initiating the process for pump replacement with a recertified pump followed by a return of the defective product. The Confirmation Form can be faxed to 1-800-628-6322 in the U.S. or 1-651-628-7485 outside the U.S., completed at www.cozmore.com/notice or calling 1-800-501-5748 in the U.S.

Distribution

Worldwide Distribution --- Including states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MI, MA, MD, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and DC and countries of AUSTRALIA, NEW ZEALAND, UNITED KINGDOM, IRELAND, CANADA, CHINA, GREECE, POLAND, SLOVENIA, SWEDEN, ISRAEL and BRAZIL.

Quantity

586