FDA Recall Terminated

Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Recall: Z-0988-2019 · Initiated January 8, 2019

Recall

Recall Number
Z-0988-2019
Event Number
82074
Firm
GETINGE US SALES LLC
FEI Number
3012092534
Product Code
FSY
Status
Terminated
Root Cause
Other
Initiated
January 8, 2019
Posted
March 7, 2019
Terminated
June 3, 2020
Address
45 Barbour Pond Dr, Wayne, NJ, 07470-2094

Description

Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.

Reason

The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.

Action

Urgent Medical Device Field Correction notification letters dated 1/8/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: -Complete and sign the enclosed Customer Response Form (Pages 4 and 5) and return to Getinge using the fax number or email address provided on the form. -A Getinge Service person will contact the person listed on the Customer Response Form (page 5) to schedule service for an update of your device, free of charge. You can also arrange a visit by a Getinge-trained or authorized service technician by calling Getinge Customer Support at (800) 950-9912 (select option 1).

Distribution

The products were distributed to the following US states: AZ and TX.

Quantity

6