FDA Recall Terminated

ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

Recall: Z-0984-2015 · Initiated November 19, 2014

Recall

Recall Number
Z-0984-2015
Event Number
69880
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
November 19, 2014
Posted
March 18, 2015
Terminated
April 28, 2015
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.

Reason

There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue.

Action

Siemens Planned Action: 1. Providing recommendations for user actions in terms of routine dosimetric quality control checks for any IMRT treatment at very high dose rates (Multiple X) as well as for mARC treatments. 2. Software update for the Function Controller #3, which controls parts of the beam generation. This software update enables the correct switching between Pulse Forming Network (PFN) and reduced Pulse Forming Network (iPFN) in pause state, to reduce the dark current during the movement as much as possible. This software will be installed by a Siemens service representative. 3. Siemens will contact the user to make an appointment to implement the modification system at a date and time when the system will be available. These corrections shall be made free of charge. The deployment of the permanent correction shall be completed by July 1, 2015. For further questions please call (610) 219-6300.

Distribution

US Distribution to the states of UT, WI, and NY.

Quantity

3